Packaging & Stability / CASE STUDY

Cosmetic wand-package testing: temperature and altitude

ARE Labs evaluated an anonymized cosmetic serum wand package in two separate experiments: a 226-hour climate-storage study and a simulated-altitude exposure study.

PackagingEnvironmental simulationProduct evaluation
Packaging and stability

How temperature and reduced pressure affected the package

Abstract

The climate experiment measured formulation and component behavior across five temperatures. A separate altitude experiment evaluated leakage and package observations at three simulated elevations. The main package-function findings occurred at elevated temperature.

Study question

The work evaluated three anonymized package variants, separating temperature-related formulation and applicator behavior from leakage risk under reduced pressure.

Selected findings

  • At 40 °C, maximum losses were 93 mg, 52 mg, and 51 mg; the 93 mg result came from one anomalous unit, while the maximum among the remaining Variant A units was 46 mg.
  • At 60 °C, all 42 evaluated units had bent wands and wipers that were displaced or stuck during opening.
  • Across 54 altitude-study units, no exterior or thread leakage was observed, while an internal-visual source inconsistency limits a broader pass claim.
Climate experiment
210 unitsThree variants across five temperatures in a study spanning 226 hours
Altitude experiment
54 unitsThree variants across 18,000, 26,000, and 40,000 ft simulated elevation
Main package signal
60 °CAll 42 units had bent wands and displaced or stuck wipers during opening

Two experiments, two development questions

Climate storage and pressure exposure answer different questions. The climate experiment assessed mass change, formulation appearance, applicator condition, and package function. The altitude experiment examined whether reduced pressure produced leakage or measurable package change. Their outcomes should not be merged into a single pass/fail claim.

TABLE 1The climate and altitude experiments were separate studies.
ExperimentUnitsConditionsObservation planPrimary question
Climate storage210 total; 70 per variant-30, 5, 23, 40, and 60 °C; study spanning 226 hoursSeven timepoints; immediate and 4-hour room-temperature observationsMass change, formulation appearance, applicator condition, and package function
Simulated altitude54 total; 18 per variant18,000, 26,000, and 40,000 ft simulated elevationImmediate and 4-hour room-temperature/pressure observationsLeakage and package behavior under reduced pressure

The experiments used the same three anonymized variant labels but were not one combined protocol.

Climate-storage design

The climate experiment included 210 units: 70 per anonymized variant, with 14 per variant at each of -30 °C, 5 °C, 23 °C, 40 °C, and 60 °C. The study spanned 226 hours. At seven observation points, two units per variant and condition were removed. One was evaluated immediately; the other was evaluated after four hours at room temperature.

The study spanned 226 hours, but individual units had different exposure durations because they were removed at successive observation points. The immediate and four-hour evaluations used separate units, so they show package appearance under each evaluation condition without demonstrating recovery of the same package between inspections. The total sample count therefore does not represent repeated measurements on every package.

Units were weighed before and after exposure and inspected for residue, internal appearance, cap and wand behavior, brush appearance, product accumulation, crystallization, and wiper seating. The qualitative accumulation scale was intentionally strict, so a slight level-1 observation did not necessarily represent an unacceptable condition.

Climate-storage results

FIGURE 1Maximum measured weight loss by storage temperatureThree anonymized variants; 14 units per variant at each temperatureGrouped bar chart of maximum measured weight loss for Variants A, B, and C at -30, 5, 23, 40, and 60 degrees Celsius.
View figure data as a table
temperatureVariant AVariant BVariant C
-30 °C0 mg0 mg0 mg
5 °C0 mg0 mg0 mg
23 °C15 mg15 mg15 mg
40 °C93 mg52 mg51 mg
60 °C88 mg95 mg90 mg

Values summarize the climate-storage experiment.

  • The 40 °C Variant A bar includes one anomalous 93 mg result; the maximum among the remaining Variant A units was 46 mg.
  • The report did not determine the cause of elevated-temperature mass loss.
TABLE 2Maximum measured weight loss in the climate-storage experiment.
TemperatureUnits across variantsVariant AVariant BVariant CObserved package/formulation condition
-30 °C420 mg0 mg0 mgFrozen immediately after removal; thawed after four hours
5 °C420 mg0 mg0 mgGenerally normal; occasional slight accumulation
23 °C4215 mg15 mg15 mgWeight loss observed; occasional accumulation
40 °C4293 mg (46 mg among remaining units)52 mg51 mgMore visible accumulation; a pre-existing residue condition dried or crystallized
60 °C4288 mg95 mg90 mgAll 42 wands bent; all wipers displaced or stuck during opening

The 93 mg result came from one anomalous Variant A unit. The maximum among the other Variant A units at 40 °C was 46 mg. Cause of mass loss was not determined.

Three abnormalities were documented before climate exposure. One Variant A unit was recorded as having an internal leak, and another Variant A unit had serum on the exterior, thread, and wand. One Variant C unit was difficult to open. The unit later photographed with dried or crystallized thread residue at 40 °C already had residue before exposure. That image therefore does not establish a new storage-caused leak.

The weight-loss display reports maxima, not group means or the fraction of units affected. At 23 °C, each variant had a maximum of 15 mg. At 60 °C, the maxima were 88, 95, and 90 mg for A, B, and C. These maxima do not show that every unit lost that amount or establish a precise temperature threshold for the package.

Maximum measured loss was 0 mg at -30 °C and 5 °C, but the immediately inspected -30 °C samples were frozen. Separate observations after four hours at room temperature recorded thawing. A zero in the weight-loss table means that this measurement recorded no mass loss under those conditions. It does not mean the product had no visible or functional change during cold exposure; the inspections describe its observed state.

The report discussed possible mechanisms for elevated-temperature mass loss, but the measurements did not establish a cause.

At 40 °C, the 93 mg result came from one unusual Variant A unit; the maximum among the remaining Variant A units was 46 mg. Both values are reported to keep that distinction clear. The pre-exposure residue observation also matters: the later photograph documents the residue's appearance but does not establish when or why it first developed.

Altitude-exposure design

The altitude experiment included 54 units: 18 per variant, with six per variant exposed at each simulated elevation. For each variant/elevation combination, three units were inspected immediately and three after four hours at room temperature and pressure. Samples were held close to horizontal in a sealed chamber for 15 minutes at the target condition.

TABLE 3Pressure and timing for the separate altitude-exposure experiment.
Simulated altitudeEquivalent absolute pressureGauge vacuumRampDwellUnits
18,000 ft7.6 psi-15.0 in Hg9 min15 min18
26,000 ft5.2 psi-19.5 in Hg13 min15 min18
40,000 ft2.7 psi-24.5 in Hg20 min15 min18
TABLE 4Mean weight change and sample standard deviation, six units per variant/elevation combination.
Variant18,000 ft26,000 ft40,000 ft
A0.167 ± 0.408 mg0.333 ± 0.516 mg0.833 ± 0.753 mg
B0.667 ± 0.816 mg0.333 ± 0.516 mg0.500 ± 0.548 mg
C0.500 ± 0.548 mg0.667 ± 0.516 mg0.833 ± 0.753 mg

The report recorded no observed exterior or cap-thread leakage. An internal-visual inconsistency between summary and inspection sheets remains disclosed in the article.

What the altitude results support

The report recorded no observed exterior or cap-thread leakage at either post-exposure time. Mean weight changes were below 1 mg for every variant/elevation combination. The internal-visual records conflict: individual inspection sheets marked the bottom viewing window as cloudy throughout, while the summary reports a moving or visible bubble. This inconsistency limits the conclusion to the supported leakage and inspection findings. The records do not support an unqualified statement that every internal-visual observation agreed.

The altitude table reports mean weight change and sample standard deviation for six units per variant and elevation. A mean below 1 mg summarizes the group; it does not mean every individual result was identical. The no-observed-leak result also does not resolve the conflicting internal-visual records. Leakage, measured mass change, and internal appearance are separate observations, each with its own limits.

Development interpretation

For this anonymized product-package configuration, elevated temperature produced more pronounced package-function findings than the tested reduced-pressure exposures. The main finding involved the applicator system: at 60 °C, wand deformation coincided with wiper displacement during opening.

High-temperature inspection identified a wand-and-wiper issue to investigate without establishing its cause. The altitude experiment provided a no-observed-leak result within the tested conditions. Neither experiment identifies a final redesign, verifies a corrective action, or establishes a shelf life. Those decisions require their own acceptance criteria and follow-up evidence; the displayed summary values alone cannot establish them.

Summary

The two experiments gave the client a comparison of package behavior under the tested temperature and reduced-pressure conditions. ARE Labs documented the observed mass changes and package-function findings, including the anomaly and conflicting visual records. These results could inform the next package review without establishing a failure mechanism or performance beyond the tested conditions.

STUDY DOWNLOADS

Take a closer look.

Explore the methods, findings and supporting detail.

01 / THE QUICK READ

Two-page summary

The study question, methods, selected results and interpretation in a concise brief.

PDF · 2 pages · 172 KB · revision 1
02 / THE TECHNICAL DETAIL

Detailed study report

A separately reviewed public edition with supporting methods, figures and result tables.

Altitude exposure technical edition · PDF · 12 pages · 938 KB · revision 1View report Download
Climate storage technical edition · PDF · 13 pages · 1.5 MB · revision 1View report Download
PUBLIC SOURCES

References and study evidence

  1. 1ASTM D6653/D6653M-13(2021): Standard Test Methods for Determining the Effects of High Altitude on Packaging Systems by Vacuum MethodASTM InternationalPublic standards context for vacuum-method evaluation of high-altitude effects on packaging systems.Open source
  2. 2Cosmetic ProductsU.S. Food and Drug AdministrationPublic category context for cosmetic products; not used as approval evidence.Open source
QUESTIONS

What to know about this study

Why were climate storage and altitude exposure evaluated separately?

Climate storage examined mass change, formulation appearance, applicator condition, and package function. Altitude exposure examined leakage and package behavior under reduced pressure. Keeping the experiments separate makes those findings easier to interpret.

What was the strongest observed package-function result?

At 60 °C, all 42 evaluated units had bent wands and wipers that were displaced or stuck during opening.

Did the altitude experiment show leakage?

No exterior or cap-thread leakage was observed in the 54 tested units at the two post-exposure observation times. The internal-visual records contained an inconsistency, which is disclosed. The conclusion is limited to the supported observations.