Purpose & when to use

ARE Labs generates precisely defined challenge atmospheres, from part-per-billion (ppb) VOC traces to percent-level oxygen-deficient blends. We use mass-flow controllers (MFCs), permeation tubes, certified gas cylinders, and Fourier-transform infrared (FTIR) spectroscopy to verify concentrations in real time. Methods align with ISO 6145 dynamic gas mixture preparation and EPA TO-15 VOC analytical frameworks under our ISO 17025 quality system. Use this service when:

  1. Validating gas detectors and air-quality sensors — ISO 6145 MFC-diluted challenge concentrations confirm detector linearity, setpoint accuracy, and dynamic response time across gas species and concentration levels.
  2. Generating repeatable VOC challenge atmospheres for [gas-phase filtration and VOC destruction studies](/testing-services/gas-voc/gas-voc-destruction/) — stable ppb-to-ppm delivery with EPA TO-15 VOC analytical framing.
  3. Supporting scrubber and ventilation system performance studies — duct-injected gas challenges with upstream and interface sampling confirm delivered concentration under ASTM D5466 canister framing.
  4. Establishing traceable ozone and by-product challenge conditions for air-cleaning device evaluations — FTIR-verified setpoints under ISO 6145 dynamic dilution support [VOC and by-product emissions reporting](/testing-services/gas-voc/voc-by-product-emissions/).
  5. Providing calibration atmospheres and step-change/ramp profiles for exposure studies — ISO 6145 dynamic dilution with documented uncertainty contributors satisfies method validation and comparability requirements.

Use gas delivery and control when your test depends on a known, stable, traceable gas concentration. Applications range from single-gas detector calibration and sensor linearity work to multi-gas chamber challenges for device performance and safety evaluations.

Gas detection, purification, and air-treatment device families

Gas delivery supports device categories wherever a controlled gas challenge is the test input — from ppb-level sensor calibration and percent-level scrubber loading to whole-room air cleaner gas-phase performance under ISO 17025 and EPA TO-15 aligned conditions.

  • Gas detectorFixed-point and portable gas detectors
  • Air quality sensorIndoor and ambient VOC / CO2 monitors
  • In-duct air cleanerDucted HVAC gas-phase modules
  • Air purifierPortable and whole-room gas-phase units
  • Breath analyzerBreath-sample gas-concentration devices

Instrumentation & measurement ranges

We select the platform based on the target gas species, required concentration range, and decision the data must support. The selection is documented during study planning and in the report.

1 – 50000 mL/minvolumetric

Mass-flow controller (MFC) stack

Primary dilution and delivery platform — calibrated MFCs blend certified gas sources to target setpoints with documented flow accuracy and dilution ratios traceable to the cylinder source.

0.1 – 1000 ppmconcentration

Gas FTIR spectrometer

Real-time multi-species concentration verification — continuously monitors delivered concentration against setpoint, documents stability bands, and captures step-change and ramp profiles.

0.001 – 10 ppmpermeation-rate

Permeation tube oven

Temperature-controlled source for low-concentration VOC and reactive gas delivery — permeation rate characterized and logged; steady-state delivery is stable over extended run durations.

1 ppm – 100%certified-concentration

Certified reference gas cylinders

NIST-traceable certified gas mixtures as primary challenge sources — cylinder certificates, lot numbers, and expiry dates logged per campaign for full source traceability.

Test method options

MethodStrengthsTradeoffAligned with
Dynamic dilution challenge (ISO 6145 aligned)
  • ISO 6145 MFC dilution delivers traceable ppb-to-ppm concentrations — setpoint accuracy and dilution uncertainty documented for detector calibration and method validation.
  • Continuous FTIR verification confirms concentration stability in real time — drift and step-change behavior captured at the measurement interface.
Line conditioning and manifold equilibration required before stable setpoints are confirmed — fast-ramp profiles need MFC command and FTIR response synchronized.
ISO 6145ISO 17025
Permeation-based challenge generation (fit for purpose)
  • Temperature-controlled permeation tubes deliver stable, low-concentration challenges for reactive gases — suited to extended runs where cylinder-based ppb precision is insufficient.
  • Permeation rate characterized before the campaign; FTIR confirms steady-state delivery before device exposure begins, providing a traceable challenge baseline.
Permeation characterization and temperature stabilization add preparation lead time — not suited to fast concentration changes or step-change response tests.—
Cylinder-blend delivery and step-change profiling (ASTM D5466 aligned)
  • Certified cylinders provide fast, robust delivery — canister-compatible sampling per ASTM D5466 supports VOC speciation and atmospheric analysis reporting.
  • Step-change and ramp profiles validate dynamic detector response under FTIR monitoring — transient concentrations documented alongside steady-state results.
Cylinder composition and dilution accuracy limit the achievable range — very low ppb targets and uncommon species use permeation delivery instead.
ASTM D5466EPA TO-15
Interface verification at device inlet (fit for purpose)
  • Samples concentration at the device inlet, not the generator outlet — documents line losses and adsorption at the point governing device performance.
  • Multi-point sampling across generator, manifold, and inlet produces a delivery chain mass balance with documented line-loss contributors for traceability.
Additional sampling ports at each measurement point increase setup complexity and run time versus single-point generator-outlet verification.—

Setup configurations

We configure each gas delivery study around the gas species, target concentration range, device interface geometry, and decision the data must support. Delivery accuracy depends on the defined dilution manifold, calibrated MFCs, and line material. These choices must also accommodate environmental controls and the device inlet geometry. The dimensions below are set during study planning:

Device interfaces

Manifold outlet, sampling port, and device inlet geometry matched to the device under test — line material (Teflon, stainless, or inert-coated) selected to minimize adsorption losses for the target gas species.

Flow & actuation profiles

Steady-state setpoint holds, step changes, and concentration ramps defined in the protocol — MFC command and FTIR response are synchronized so dynamic behavior at the device inlet is fully documented.

Environmental controls

RH and temperature setpoints and stabilization periods defined before delivery begins — conditioning reduces gas-line artifacts and improves setpoint stability for humidity-sensitive species.

Sample numbers

Replicate setpoint holds and repeat sequences confirm delivery repeatability — replicate plan sized to expected variability from MFC accuracy, permeation rate stability, and FTIR noise floor.

Calibration & verification

Pre- and post-run MFC flow verification and FTIR baseline checks — leak checks on manifold connections and line-loss assessment at representative sampling points before challenge delivery begins.

Methods anchored to the standards that matter

Gas delivery studies follow a documented quality system anchored to the gas mixture preparation, VOC analytical, and laboratory competence frames that govern traceable challenge atmosphere generation. The four anchors below carry through to §7 outputs.

  • ISO 17025AccreditedTesting-laboratory competence — documented methods, calibration traceability, and uncertainty contributors.
  • ISO 6145AlignedDynamic preparation of calibration gas mixtures — MFC dilution accuracy and uncertainty documentation.
  • EPA TO-15AlignedVOC determination in ambient air — canister sampling and GC/MS analytical framing for VOC speciation.
  • ASTM D5466AlignedAtmospheric VOC canister sampling — collection and chain-of-custody practices for gas-phase VOC campaigns.

Key data outputs & reporting

You receive delivered-concentration time series, stability metrics, and flow conditions. The report presents primary results, QA / QC controls, and uncertainty contributors in a format for device performance files, method validation reports, or detector calibration documentation. The deliverables below cover the standard report. Extended studies comparing multiple gas species, setpoints, or device configurations include additional comparison artifacts.

Primary outputs

  • Delivered concentration versus time (setpoint vs. FTIR-measured) with mean, SD, and stability band documented per setpoint hold.
  • Step-change and ramp profiles showing concentration rise time, overshoot, and settle time at the device inlet when dynamic response is the study objective.
  • Flow conditions and dilution ratios for each setpoint — MFC settings, source concentration, and dilution factor logged for traceability.
  • Line-loss and delivery-chain mass balance where multi-point sampling is included — generator outlet versus manifold versus device inlet concentrations compared.

Deliverables

#FormatContents
01PDF reportMethods, flow conditions, stability metrics, and uncertainty contributors per setpoint.
02CSV / XLSX datasetsTime-series concentration, setpoint tracking, and stability statistics per condition.
03FiguresSetpoint-vs-measured overlays, stability bands, and ramp/step-change profiles for technical files.
Extended deliverables · multi-arm comparability · stability · predicate studies
  • Multi-species comparison pack — Side-by-side setpoint tracking and stability overlays across gas species or concentration levels.
  • Design-change delivery delta — Before/after delivery accuracy and line-loss comparison for manifold, tube material, or interface changes.

QA / QC & data integrity

Each gas delivery study includes documented QA / QC matched to the instrument suite, gas species, and decision the data supports. Verifications run before and after each test set and are audited under our ISO 17025 quality system. Calibration records provide traceability from MFC characterization and cylinder certification through the final delivered-concentration result. Every campaign includes leak checks and line-loss assessments.

Calibration and verification logs for all mass-flow controllers — pre- and post-run flow checks against a calibrated reference confirm MFC accuracy within the study's acceptance window.

FTIR baseline validation and multi-point verification before challenge delivery — confirms instrument response and detection sensitivity for the target gas species at the study's concentration range.

Leak checks on manifold connections and sampling lines before challenge introduction — pressure-hold or gas-sniff checks confirm no unintended dilution or bypass along the delivery path.

Line-loss and adsorption evaluation at representative sampling points — documents delivery-chain losses between generator outlet and device inlet for traceability to the final result.

Chain of custody for permeation sources and certified gas cylinders — cylinder certificates, lot numbers, expiry dates, and permeation characterization records kept with the study file.

Why ARE Labs

ARE Labs connects technical topics to practical study design, method selection, controlled aerosol work, and reportable evidence without turning technical pages into sales pages.

Reviewed byJamie Balarashti (25 yrs - cascade & inhalation methods) - Weston Schaper (7 yrs - real-time sizing & nanoparticle work)
17025Accredited testing
900+Studies Performed
17+Years in operation
300+Clients supported

Common questions

These answers cover gas species coverage, concentration range, RH conditioning, dynamic profiling, and deliverables for gas-detector developers, air-quality sensor teams, scrubber engineers, and exposure-study researchers. They provide a starting point for planning a study. Contact us to discuss a target species, concentration, or device interface beyond those shown here. Most campaigns require at least one configuration choice specific to the gas and device.

Q.What gas species can you deliver and verify?
A.We generate challenges for VOCs, CO2, NOx, SOx, ozone, water vapor, oxygen-deficient blends, and other gases using permeation tubes or certified cylinders. FTIR covers a broad species range, and we add GC/MS per EPA TO-15 for speciated VOC verification.
Q.How do you control and verify delivered concentration at the device inlet?
A.MFCs control dilution from a certified source to the target setpoint. FTIR monitors delivered concentration against that setpoint in real time. For interface verification, we add a sampling port at the device inlet and compare generator-outlet and inlet concentrations to document line losses.
Q.Why is RH and temperature conditioning important for gas delivery?
A.Many gas species and delivery lines are sensitive to humidity and temperature. Conditioning the carrier gas and stabilizing line temperature before delivery begins reduces adsorption losses, improves setpoint stability, and limits measurement artifacts, especially for reactive or polar VOCs.
Q.Can you run step changes and concentration ramps for dynamic testing?
A.Yes. Step-change and ramp profiles are standard for detector response and control-system validation. We log MFC command and FTIR response with synchronized timestamps to accurately capture rise time, overshoot, and settle time at the device measurement point.
Q.How is the gas delivery chain documented for traceability?
A.We log cylinder certificates, lot numbers, and expiry dates; MFC calibration records; FTIR verification; dilution ratios; and line-loss measurements. For each setpoint condition, we report ISO 6145 uncertainty contributors: flow accuracy, mixing completeness, adsorption, and measurement noise.
Q.What do I receive at the end of the study?
A.You receive a PDF report covering methods, flow conditions, stability metrics, and uncertainty; CSV / XLSX time-series datasets; and figures showing setpoint-versus-measured overlays, stability bands, and any step-change or ramp profiles from the campaign.