Aerosol Characterization / CASE STUDY

Fragrance aerosol testing: particle size and chamber counts

ARE Labs compared an anonymized fragrance aerosol group with three comparator groups using laser diffraction and time-resolved chamber particle measurements. Results apply only to the tested products and conditions and do not establish odor longevity, safety, exposure, or clinical performance.

FragranceParticle sizingProduct evaluation
Fragrance aerosol comparison

Two methods, two distinct aerosol questions

Abstract

Laser diffraction measured volume-based spray particle-size distributions for 10 products. A fast mobility particle sizer then measured chamber counts over one hour for one representative product from each group. The two methods used different sample sets and reported different endpoints.

Study question

The study compared the tested fragrance aerosols by spray particle-size distribution and time-resolved chamber counts under defined conditions.

Selected findings

  • Group-average D[4,3] was 31.49 µm for Test Article A and 75.03-89.77 µm for Comparator groups B-D.
  • The chamber products had unequal actuation durations and substantially different starting counts.
  • Particle residence in this chamber does not establish odor longevity, safety, exposure, lung deposition, or clinical performance.
PSD products
10Three products in A, B, and C; one product in D
Chamber products
4One representative product from each anonymized group
Observation
60 minFMPS sampling at 30-second intervals
FMPS range
5.6-560 nmNumber concentration by electrical-mobility size

Study question

The study asked two questions about the aerosols' physical characteristics. How did their volume-based spray particle-size distributions compare at release? How did instrument counts change over one hour in a controlled chamber for one representative product from each group?

Phase 1: spray particle-size distribution

ARE Labs used a Malvern Spraytec laser-diffraction system to measure each product at six inches from the detection laser. Each product was hand-actuated at least four times, and the measurements were averaged. D10, D50, and D90 are volume percentiles; D[4,3] is the volume-weighted mean diameter.

The table reports size on a volume basis. D50 marks the midpoint of the measured volume distribution; it is not a median based on particle count. D[4,3] and D50 describe different aspects of that same distribution, so their values can differ without indicating disagreement between instruments. They are not two independent measures of fragrance performance.2

The groups contained unequal numbers of products. A, B, and C each included three products; D included one. The displayed values are descriptive group averages, not inferential estimates of every product in a category.

Repeated hand actuations do not increase the number of independently sampled products. Averaging at least four actuations describes the measurement procedure for each product; group D still contains only one product. The three-product group averages and the one-product comparator therefore have different sample structures and do not provide equally broad estimates of a commercial category.

TABLE 1Displayed group-average volume-based particle-size results
Product groupPSD productsD[4,3]D50
Test Article A331.49 µm28.49 µm
Comparator B375.03 µm69.72 µm
Comparator C377.91 µm69.26 µm
Comparator D189.77 µm80.64 µm

A, B, and C are averages of three products each; D is one product. Values retain the source report's displayed precision.

FIGURE 1Particle-size comparisonDisplayed group-average volume-based metrics for the tested productsGrouped bar chart comparing D[4,3] and D50 values for Test Article A and Comparator groups B, C, and D.
View figure data as a table
groupD[4,3]D50
A31.49 µm28.49 µm
B75.03 µm69.72 µm
C77.91 µm69.26 µm
D89.77 µm80.64 µm
  • A, B, and C are group averages of three products; D represents one product.
  • The chart is descriptive and does not show inferential uncertainty.

Under the reported measurement conditions, A had the lowest displayed group-average D[4,3] and D50 values. This is a result for the tested product set, not a universal ranking or evidence of sensory preference.

Phase 2: chamber count over time

One representative product from each group was dispersed into a sealed 16 m³ chamber. The report describes approximately 22 °C and 35% relative humidity, two mixing fans, and two chamber air changes per hour. A TSI Fast Mobility Particle Sizer model 3091 sampled at 30-second intervals during one-hour trials.

The study reports observations at 30-second intervals. Although the manufacturer lists a faster instrument time resolution, that capability does not establish that readings were retained or analyzed at the faster rate in this study. Interpretation should follow the reported study record.3

The FMPS measurement range was 5.6-560 nm. It measured electrical-mobility size and number concentration, but could not determine whether an individual measurement represented a particle or droplet or identify its chemical composition. The report also describes an aerodynamic particle sizer as a pretest instrument. Its role should be kept separate from the FMPS trial data and the optical Spraytec results.

The two phases used different measurement locations, size ranges, and distribution bases. A volume-based spray diameter measured near release cannot be substituted directly for a chamber number concentration in the FMPS range. The chamber counts do not track individual droplets measured by laser diffraction or describe material outside the selected instrument range. The phases address different questions, and their displayed values are not interchangeable endpoints.2,3

The chamber inputs were not equivalent. A3 was actuated for approximately 750 ms; B1, C2, and the single group D product were actuated for one second. The report does not provide emitted-mass normalization.

TABLE 2Starting FMPS counts in the 5.6-560 nm measurement range
Chamber productActuationStarting count
A3Approximately 750 ms1.62E+05 particles/cm³
B11 second4.46E+02 particles/cm³
C21 second6.71E+03 particles/cm³
D11 second2.93E+02 particles/cm³

The starting counts are not normalized to emitted mass or an equivalent delivered dose. Figure 12 in the source plots raw counts on a logarithmic scale.

The A3 count trace remained above the comparator traces throughout the reported hour. A3 also began substantially higher than the comparators, and its actuation duration differed. The report provides no normalized decay constants, half-lives, equivalent-dose comparisons, confidence intervals, or replicated chamber-trial statistics.

A higher trace throughout the hour does not, by itself, show a slower fractional decline. That comparison would need to account for initial concentration and relevant input differences. The starting-count table shows that the products entered the observation period at markedly different levels. The raw counts should therefore be reported separately from any normalized persistence claim, without assuming an unreported decay model.

Interpretation boundaries

The particle-size results support a specific comparison: A produced lower displayed volume-based size metrics than B-D in this tested sample set. The chamber results support a separate observation: the measured A3 count trace began higher and remained higher during the reported one-hour trial.

Neither result establishes how long people perceive a fragrance. Odor perception depends on chemical composition, vapor behavior, sensory thresholds, ventilation, deposition, use conditions, and other factors not measured here. Particle residence in this chamber does not establish odor longevity.

The study also does not establish inhalation exposure, lung deposition, toxicological safety, or clinical effect. The FMPS did not identify chemical composition, and the study did not measure breathing-zone exposure or biological response.

The historical report contains conflicting regulatory references in its GLP statements. It also contains multiple source dates, none of which has been established as a publication date. The report does not establish this study's accreditation or GLP status. ARE Labs' current laboratory accreditation is described in the related Quality Documentation page.

What the study contributes

Within these limits, the work shows how complementary aerosol methods answer different questions. A future claim-support program should define the intended claim first, normalize the relevant inputs, and predefine endpoints. Sensory, exposure, deposition, or safety methods should be added only when those outcomes are at issue.

Summary

For the client, the work distinguished volume-based particle sizing from one-hour chamber particle measurements. ARE Labs documented the tested product comparison and its limits: unequal inputs and starting counts prevent a normalized persistence claim. The findings provide a basis for planning a future claim-specific study; they do not establish odor longevity, safety, exposure, or clinical performance.

STUDY DOWNLOADS

Take a closer look.

Explore the methods, findings and supporting detail.

01 / THE QUICK READ

Two-page summary

The study question, methods, selected results and interpretation in a concise brief.

PDF · 2 pages · 173 KB · revision 1
02 / THE TECHNICAL DETAIL

Detailed study report

A separately reviewed public edition with supporting methods, figures and result tables.

Fragrance aerosol technical edition · PDF · 28 pages · 2.3 MB · revision 1View report Download
PUBLIC SOURCES

References and study evidence

  1. 1ISO 13320:2020, Particle size analysis - Laser diffraction methodsInternational Organization for StandardizationPublic standards context for laser diffraction particle size analysis.Open source
  2. 2Definition and measurement methods of particle size distributionMalvern PanalyticalOpen source
  3. 3Fast Mobility Particle Sizer (FMPS) 3091TSI IncorporatedOpen source
QUESTIONS

What to know about this study

Does the smaller measured spray size prove the fragrance lasts longer?

No. The study did not measure perceived odor duration. Particle size and raw chamber counts cannot establish sensory longevity by themselves.

Did the chamber phase compare equivalent doses?

No equivalent-dose normalization is reported. The test article and comparators had different actuation durations and substantially different starting counts.

Are the chamber counts a safety or exposure result?

No. They are instrument counts in a defined size range under chamber conditions. They are not a breathing-zone exposure estimate, chemical analysis, toxicology result, or clinical endpoint.

Why are the products anonymized?

The client, brands, product names, source filenames, and traceable sample details remain private. Controlled aliases preserve the comparison structure without disclosing identities.