Aerosol Characterization / CASE STUDY

Data Center Particle Deposition: Surface Contamination

A week-long study across two data center buildings measured where particles settled, how much mass accumulated, and what the deposited particles looked like under optical microscopy.

Data centerParticle depositionField studyParticle sizing
Data center case study

Surface deposition: building averages and local differences

Abstract

ARE Labs evaluated deposited particles in two data center buildings, each with four halls. Collection slides sampled three position types: floors, equipment-height goalposts, and overhead plenums.

Study question

The client needed a measured baseline for surface particle accumulation across operating spaces, with enough spatial detail to compare buildings, halls, and sampling heights.

Selected findings

  • Across 123 primary slides, mean and median mass deposition rates were 14.7 and 7.5 mg/m²/day.
  • Goalpost positions had the highest building-level mass means, while plenum positions averaged 5.5 mg/m²/day in both buildings.
  • Among 12,501 retained microscopy objects, 76% were smaller than 2 µm, 93% smaller than 3 µm, and 98% smaller than 5 µm.
Coverage
2 buildingsFour data halls per building, with floor, goalpost, and plenum collection positions.
Primary dataset
123 slides126 slides were deployed; three were excluded from the primary analysis under the report criteria.
Exposure period
7–8 daysCollection slides remained in the operating facilities for approximately one week.

The operational question

The client needed to understand where particles accumulated within the data center. Surface deposition measurements show what reaches equipment-height structures, floors, and overhead spaces over time. The baseline needed to capture local accumulation patterns that a single facility-wide average could obscure.

The study covered two operating buildings, each with four data halls. Collection slides were positioned on floors, on equipment-height goalposts, and in overhead plenums. Using the same three position types in each hall allowed comparisons across all eight halls.

A field collection and laboratory analysis workflow

ARE Labs supplied the collection slides and deployment protocol. The client placed and recovered the slides at the facilities, then returned them to ARE Labs for laboratory analysis.

The client deployed 126 slides for approximately one week. Three slides were excluded from the primary dataset under the report criteria, leaving 123 records for the main mass-deposition analysis. ARE Labs divided the signed mass change by collection area and deployment duration to calculate a daily surface deposition rate in mg/m²/day.

The full preparation set contained 135 slides: 126 deployed slides and nine reference specimens. The reference specimens included three unopened shipping controls, four blanks, and two unused slides. They were prepared and measured on the same schedule but were never exposed at the facility. Their results are therefore reported as mass change and counts per slide rather than deposition rates.

Each slide was weighed twice before shipment and twice after return. ARE Labs averaged the two readings from each stage, then divided the signed mass change by the nominal 3 × 1 inch slide area of 0.00193548 m² and the recorded exposure duration. Negative changes were retained as measured. This calculation put slides with different seven- to eight-day exposure periods on the same daily surface-area basis.

The analysis combined gravimetry with two optical assessments. Whole-slide inspection counted a distinct coarse population nominally larger than 100 µm. Microscopy and image analysis characterized the much smaller retained objects. These populations answer different questions, so their results are reported separately rather than combined into one particle count.

TABLE 1Aggregate surface-deposition and microscopy results.
MeasurementValid recordsReported result
Mass deposition rate123 slidesMean 14.7; median 7.5 mg/m²/day
Coarse whole-slide count121 slidesMean 2,090; median 1,774 particles/m²/day
Microscopy size analysis12,501 objects76% <2 µm; 93% <3 µm; 98% <5 µm

Gross results are shown without background subtraction. Whole-slide coarse counts and retained microscopy objects are distinct optical populations.

Deposition mass varied by sampling position

Across the 123-slide primary dataset, the mean mass deposition rate was 14.7 mg/m²/day and the median was 7.5 mg/m²/day. Some slides accumulated substantially more material than the typical slide, raising the overall mean. The gap between mean and median shows why a single average cannot describe every location.

Mean mass deposition rates varied by position within each building. In Building 1, floor, goalpost, and plenum locations averaged 17.6, 18.3, and 5.5 mg/m²/day. Building 2 averaged 12.9, 20.8, and 5.5 mg/m²/day at the same positions. Goalposts had the highest position mean in both buildings; plenums had the lowest, with the same mean in each building.

FIGURE 1Total particle mass per area across one data hall over the seven-day collection (mg/m²). Contours interpolate between sampling locations; the color scale applies to this hall only.
Anonymized Hall A heat map of total particle mass per area, with cooler colors at the entrance and warmer colors toward the rear.

The heat map shows local differences in accumulated mass within one hall that a hall-wide average could conceal. Its mg/m² scale represents the full collection period, while the adjacent comparison chart reports daily rates in mg/m²/day. Marked values identify the collection locations. Colors between them show an interpolated pattern that can help guide follow-up sampling; they do not represent measurements at every point.

FIGURE 2Mean mass deposition rate by building and positionGoalpost positions had the highest building-level means; plenum means were lower in both buildings.Grouped bar chart comparing mean mass deposition rates at floor, goalpost, and plenum positions in two anonymized data center buildings.
View figure data as a table
positionBuilding 1Building 2
Floor17.6mg/m²/day12.9mg/m²/day
Goalpost18.3mg/m²/day20.8mg/m²/day
Plenum5.5mg/m²/day5.5mg/m²/day

Source: anonymized ARE Labs data center particle deposition report.

  • Values are gross mass deposition rates without background subtraction.
  • Building and hall identities are anonymized.

These location differences do not identify the source or transport mechanism of the deposited material. They do show where follow-up investigation or repeated monitoring may be most useful. A facility can use the comparisons to choose future collection locations or decide where to examine maintenance and traffic patterns or collect a longer baseline.

Optical analysis resolved two particle populations

The coarse whole-slide count used 121 valid slides and produced a mean of 2,090 particles/m²/day and a median of 1,774 particles/m²/day. This count represents the report's nominally greater-than-100-µm optical population. Whole-slide inspection and microscopy used different detection windows, so the coarse count should not be added to or directly compared with the microscopy object count.

Microscopy retained 12,501 objects for size analysis. Of those objects, 76% were smaller than 2 µm, 93% smaller than 3 µm, and 98% smaller than 5 µm. The microscopy workflow examined the 1.12-100 µm range under two magnifications. Particles larger than 100 µm were counted by hand on whole slides but were not sized.

Calibrated microscopy and image analysis characterized particles in selected fields. The reported size profile describes the examined fields rather than every particle on a slide.1

FIGURE 3Cumulative size distribution of retained microscopy objectsMost retained objects measured below 5 µm in the optical microscopy analysis.Bar chart showing that 76 percent of retained microscopy objects were smaller than 2 micrometers, 93 percent smaller than 3 micrometers, and 98 percent smaller than 5 micrometers.
View figure data as a table
thresholdCumulative share
<2 µm76% of retained objects
<3 µm93% of retained objects
<5 µm98% of retained objects

Source: anonymized ARE Labs data center particle deposition report.

  • Percentages are cumulative within the 12,501 retained microscopy objects.

What the results support

The study established a surface-deposition baseline across two buildings, using the same three sampling heights throughout. The results allow mass accumulation to be compared by location and separately describe coarse whole-slide counts and the size distribution of retained microscopy objects. The report's heat maps show how deposition varied across the sampled halls.

The reported values are gross deposition results without background subtraction. The study did not determine particle source or composition, measure airborne concentration, or assign an ISO cleanliness class. Those questions require additional sampling or analytical work. Optical microscopy describes size and morphology; SEM-EDS can be added when elemental composition is part of the project question.

Summary

The week-long survey gave the client a measured baseline for comparing buildings and sampling heights, with 123 primary slides in the mass-deposition analysis. Separate optical analyses described coarse deposited particles and the size distribution of retained microscopy objects. ARE Labs supplied the collection media and protocol, performed the laboratory analysis, and reported the findings to support monitoring and follow-up decisions.

STUDY DOWNLOADS

Take a closer look.

Explore the methods, findings and supporting detail.

02 / THE TECHNICAL DETAIL

Detailed study report

A separately reviewed public edition with supporting methods, figures and result tables.

Data center particle deposition study report · PDF · 18 pages · 2.5 MB · revision 2View report Download
PUBLIC SOURCES

References and study evidence

  1. 1CellProfiler 4: improvements in speed, utility and usabilityBMC BioinformaticsPrimary publication describing the image-analysis platform used for automated object identification and measurement.Open source
QUESTIONS

What to know about this study

What did the data center particle deposition study measure?

The study measured surface deposition mass, a coarse whole-slide optical count, and the size distribution of retained microscopy objects across floor, goalpost, and plenum positions.

Did ARE Labs deploy the collection slides onsite?

ARE Labs supplied the slides and collection protocol. The client deployed and recovered the slides, then returned them to ARE Labs for laboratory analysis.

Did the study identify what the particles were made of?

No. The study measured deposited mass and optical particle populations. Source attribution and elemental composition were outside its scope; SEM-EDS can be added when composition is a project objective.

Can these results be used as an ISO cleanliness class?

No. The results describe deposited material on collection slides, not airborne particle concentration, so the study did not assign an ISO cleanliness class.